Using the Turbidimeter
The turbidimeter measures the light transmittance of a sample in NTU's
(Nephelometric Turbidity Units, a standard measure). It needs no
field calibration. Handle the sample vials only by their ends
(preferably the lid) so as not to affect the transmittance; wipe any
fingerprints, spots, etc. from the outside of the vial; and be sure to
close the vial-compartment lid when taking a measurement. The
turbidimeter should display ``AUTO RNG'' (for auto-range selection)
and ``SIG AVG'' (for take an average reading) when ready for use
(Fig. 4.7).
Figure 4.7:
Turbidimeter controls and
display. Press ``READ'' when ready to take a measurement. After Hach
Model 2100P Turbidimeter Instruction Manual, 1997, p. 24.
|
Use the following procedure when measuring the turbidity of your
sample:
- Fill turbidity vial (has white line around top of glass with
downward arrow) to the line (about 15 mL) with
unfiltered water. Take care to handle the sample cell by
the top. Cap the cell.
- Wipe the cell with a soft, lint-free cloth to remove water spots and
fingerprints.
- Press I/O - the instrument will turn on. Place the instrument on a
flat surface. Do not hold the instrument while making readings.
- Put the sample vial in the instrument cell compartment so its
diamond mark aligns with the raised orientation mark in
front of the cell compartment. Close the cover.
- Set automatic range by pressing the RANGE key. The display will
show ``AUTO RNG''.
- Select signal averaging (reports average of 10 measurements)
by pressing ``Signal Average'' key, display should show ``SIG AVG''
- Press READ - the display will show ``--NTU'' and the light
bulb icon will flash 10 times (once for each reading). The final
average will be displayed as numbers in NTU after the lamp symbol
turns off.